Detection of Pulmonary Metastases with Multi–Detector Row CT Scans of 5-mm Nominal Section Thickness: Autopsy Lung Study

PURPOSE: To determine the effect of changing pitch and collimation on depiction of pulmonary metastases on scans of 5-mm section thickness obtained with multi–detector row computed tomography (CT) compared with those obtained with single–detector row CT.

MATERIALS AND METHODS: In five autopsy lungs, 1,013 metastatic 0.5–30.0-mm nodules were detected at helical CT with 1-mm collimation and histopathologically diagnosed as metastases. Each nodule was numbered, and its localization was recorded as the standard for subsequent studies. Four types of scans of 5-mm section thickness were obtained with multi–detector row CT and four sets of helical pitch and table speed, respectively, as follows: set A, 3:1 and 7.5 mm per rotation; set B, 6:1 and 15 mm per rotation; set C, 6:1 and 30 mm per rotation; set D, conventional and 5-mm interval. Conventional helical CT scans (set E) were obtained with 5-mm collimation at single–detector row CT. Two independent observers evaluated the five sets of CT scans.

RESULTS: Acquisition times for sets A–D, respectively, were 1.9, 3.8, 7.5, and 1.5 times faster than they were for set E. The mean numbers of detected nodules were 671 (66%) in set A, 661 (65%) in set B, 678 (67%) in set C, 654 (65%) in set D, and 656 (65%) in set E; there was no significant difference in the number of detected nodules among the five sets (P = .997, McNemar test and Bonferroni equation).

CONCLUSION: Regardless of varying pitch or detector collimation, multi– and single–detector row CT scans obtained with 5-mm section thickness have almost the same ability to depict pulmonary metastases and are equivalent.

© RSNA, 2002

References

  • 1 Rydberg J, Buckwalter KA, Caldemeyer KS, et al. Multisection CT: scanning techniques and clinical applications. RadioGraphics 2000; 20:1787-1806. LinkGoogle Scholar
  • 2 Klingenbeck-Regn K, Schaller S, Flohr T, Ohnesorge B, Kopp AF, Baum U. Subsecond multi-slice computed tomography: basics and applications. Eur J Radiol 1999; 31:110-124. Crossref, MedlineGoogle Scholar
  • 3 Muhm JR, Brown LR, Crowe JK, Sheedy PF, Hattery PR, Stephens DH. Comparison of whole lung tomography and computed tomography for detecting pulmonary nodules. AJR Am J Roentgenol 1978; 131:981-984. Crossref, MedlineGoogle Scholar
  • 4 Chang AE, Schaner EG, Conkle DM, Flye MW, Doppman JL, Rosenberg SA. Evaluation of computed tomography in the detection of pulmonary metastases: a prospective study. Cancer 1979; 43:913-916. Crossref, MedlineGoogle Scholar
  • 5 Krudy AG, Doppman JL, Herdt JR. Failure to detect a 1.5 centimeter lung nodule by chest computed tomography. J Comput Assist Tomogr 1982; 6:1178-1180. Crossref, MedlineGoogle Scholar
  • 6 Vock P, Soucek M, Daepp M, Kalender WA. Lung: spiral volumetric CT with single-breath-hold technique. Radiology 1990; 176:864-867. LinkGoogle Scholar
  • 7 Costello P, Anderson W, Blume D. Pulmonary nodule: evaluation with spiral volumetric CT. Radiology 1991; 179:875-876. LinkGoogle Scholar
  • 8 Friese SA, Rieber A, Fleiter T, Brambs HJ, Claussen CD. Pulmonary nodules in spiral volumetric and single slice computed tomography. Eur J Radiol 1994; 18:48-51. Crossref, MedlineGoogle Scholar
  • 9 Remy-Jardin M, Remy J, Giraud F, Marquette CH. Pulmonary nodules: detection with thick-section spiral CT versus conventional CT. Radiology 1993; 187:513-520. LinkGoogle Scholar
  • 10 Markardian B, Daily ET. Preparation of inflated lung specimens. In: Heitzman ER, eds. The lung: radiologic-pathologic correlations. 2nd ed. St Louis, Mo: Mosby, 1984; 4-12. Google Scholar
  • 11 Polacin A, Kalender WA, Marchal G. Evaluation of section sensitivity profiles and image noise in spiral CT. Radiology 1992; 185:29-35. LinkGoogle Scholar
  • 12 Diederich S, Lentschig MG, Winter F, Roos N, Bongartz G. Detection of pulmonary nodules with overlapping vs non-overlapping image reconstruction at spiral CT. Eur Radiol 1999; 9:281-286. Crossref, MedlineGoogle Scholar
  • 13 Diederich S, Semik M, Lentschig MG, et al. Helical CT of pulmonary nodules in patients with extrathoracic malignancy: CT-surgical correlation. AJR Am J Roentgenol 1999; 172:353-360. Crossref, MedlineGoogle Scholar
  • 14 McCollough CH, Zink FE. Performance evaluation of a multi-slice CT system. Med Phys 1999; 26:2223-2230. Crossref, MedlineGoogle Scholar
  • 15 Hu H, He HD, Foley WD, Fox SH. Four multidetector-row helical CT: image quality and volume coverage speed. Radiology 2000; 215:55-62. LinkGoogle Scholar

Article History

Published in print: Jan 2003